Calibration vs. Verification

Calibration states a measured relationship with an uncertainty. Performance verification, run against ISO 10360-2 or ASME B89.4.1, confirms the machine still meets its specification. Most work ordered as "CMM calibration" is verification, often paired with geometric adjustment.

Where the controller allows it, the volumetric compensation map is updated during that adjustment. Some controllers are OEM-locked and compensation access is limited to the OEM's own service channel.

What a Visit Covers

Each step of a verification visit produces its own recorded result, following a fixed sequence rather than a single pass-fail check.

  • 01

    Scales and reader heads

    X, Y and Z scales are inspected and cleaned before any test proceeds further.

  • 02

    Air bearings and lines

    Bearing condition and airline supply are checked for the pressure the frame requires.

  • 03

    Geometry

    Straightness, linearity and squareness are measured in the XY, ZX and ZY planes.

  • 04

    Probing error

    Form error is tested per ISO 10360-5 against a dedicated calibrated sphere, separate from the probe's reference sphere.

Sequence of Work

  • 01

    As-found test

    Length measurement error is recorded at 105 points before any adjustment is made.

  • 02

    Adjustment

    Geometric correction and, where supported, volumetric compensation are applied.

  • 03

    As-left verification

    The same test sequence confirms the machine now meets E0,MPE.

  • 04

    Certificate issue

    Results, uncertainty and the standard applied are documented under the performing laboratory's scope.

Why St. Louis CMM Calibration and Repair

  • 01

    105-point length test

    Five calibrated lengths, three repeats, seven positions, per ISO 10360-2.

  • 02

    Repair alongside verification

    Probe heads, scales, reader heads, air bearings and drives are serviced under one visit.

  • 03

    Before-and-after reported

    Where adjustment occurs, results before and after are both stated on the certificate.